Ch2-Basic Physics of Medical Imaging.
Contents
1-X-Ray beam
attenuation
2- X-Ray tube
3-Tube rating and
cooling
4- Line focus principle
5-Anode heel effect
6-off focus radiation
7-Questions
1-X-Ray beam
attenuation:
·
X-Ray is
produced from an X-ray tube with different energies (Poly-Energetic).
·
The maximum
X-ray photon energy is equal to the pre-adjusted Kev parameter.
·
The lowest
energy of x-ray photons depends on the X-Ray tube filtration.
·
When an X-ray
incident on any material, the intensity of the x-ray is divided into three
parts. The first part is absorbed (attenuated), the second part is scattered,
and the 3rd part is transmitted.
·
The intensity of
transmitted radiation depends on the attenuation property (attenuation coefficient),
the thickness of the material, and radiation energy.
·
Higher thickness
and attenuation coefficient cause higher X-ray attenuation, and less
transmission.
·
Higher radiation
energy causes less attenuation and increases the thickness required for
attenuation.
·
The thickness
required to reduce the intensity of radiation to half is the half-value layer
HVL.
· For monoenergetic X-Ray:
·
If the X-ray
beam passes through a certain material with HVL, the intensity will be reduced
to half.
·
the transmitted X-ray
beam will have the same energy but half the intensity of the primary X-ray.
·
the reduction of
transmitted X-ray intensity to half requires the same thickness HVL, as no
difference in Energy.
·
the first HVL is
equal to the second HVL, as the mean energy of the primary x-ray is equal to
the mean energy of the transmitted X-Ray.
·
The reduction in
x-ray quantity (beam intensity) not quality (beam energy)
· For poly-energetic X-Ray:
·
if the X-ray photons carry a range of energies
starting from10 Kev to 100 Kev, so the mean energy is 50 Kev. After the first
HVL, the transmitted X-ray photons will carry energies from 50 Kev to 100Kev,so
the mean energy will be 75Kev, the mean energy (75 Kev) of the transmitted
X-Ray requires a higher thickness of the same material to be reduced to half
intensity (higher HVL).
·
For
poly-energetic X-ray: the second HVL is higher than the first HVL of the same
material (Beam Hardening).
· 2- X-Ray tube
1-Glass
enclosure
·
Necessary to
seal the two electrodes of the x-ray tube in a vacuum.
·
The vacuum →
allows the electrons to move rapidly from cathode to anode without blockage.
·
The connecting
wires must be sealed into the glass wall of the x-ray tube.
·
Special alloys, having approximately the
same coefficients of linear expansion as Pyrex
· glass, are generally used in x-ray tubes
2-Cathode Filament
• X-ray
tube has a long filament for large focal spots and short filaments for small
focal spot size
• Made
of thin (0.2 mm) tungsten wire
• has
a high atomic number (A 184, Z 74)
• is
a good thermionic emitter (good at emitting electrons)
• can
be manufactured into a thin wire that has a very high melting temperature
(3422°c)
• The
size of the filament relates to the size of the focal spot. Some
cathodes
have two filaments for broad and fine focusing.
• As
current passes through the filament, the atomic and electronic motion in metal
is sufficiently violent to enable a fraction of the free electrons to leave the
surface despite the net attractive pull of the lattice of positive ions.
• The
electrons are then repelled by the negative cathode and attracted by the
positive anode. Because of the vacuum, they are not hindered in any way and
bombard the target with a velocity around half the speed of light.
• Tungsten
is chosen for use in x-ray tubes, because:
1.
It can be drawn into a thin wire that is quite strong.
2.
Has a high melting point (3370° C).
3.
Has little tendency to vaporize.
∴ Tungsten filament has
a reasonably long-life expectancy
• When
current flows through tungsten wire → heated → its atoms absorb thermal energy→
some of the electrons acquire energy → move a small distance from the metal
surface → form a small cloud in the vicinity of the filament “the space
charge”.
• The
electron cloud, produced by thermionic emission, is also termed "Edison
effect".
• Tube
potentials (Kv) enable larger x-ray tube current for the same filament current;
for instance, for the same filament current of 5 A at 80 kV, a tube current of
800 mA is produced, whereas at 120 kV a tube current of about 1100 mA results.
SATURATION
VOLTAGE
•
If the potential applied across the tube is insufficient to cause almost all
electrons to be
pulled
away from the filament when they are emitted → Space Charge Effect: a cloud of
negative
charges tend to prevent other electrons from being emitted from the filament
until
they have acquired sufficient thermal energy to overcome the force caused by
the
space
charge → limits the number of electrons → limits X-ray tube current.
•
Below the saturation point,
the tube current is limited by the space charge effect (space-charge-limited).
↑↑ kV → significant ↑↑ in x-ray tube current although filament heating is the
same.
•
Above the saturation voltage,
the space charge effect has no influence on the x-ray tube current.
the tube current is determined by the number of electrons made available by the
heated
filament (emission-limited or temperature-limited).
↑↑ kV → very little change in tube current
3-Focusing cup
· Made
of molybdenum which has a high melting point
· poor
thermionic emitter so electrons aren't released to interfere with an electron
beam from the filament
· Negatively
charged to focus the electrons towards the anode and stop spatial spreading
Cathode focusing cup:
· Surrounds
the filament & maintained the same negative potential as the filament.
· So,
its electrical forces cause the electron stream to converge onto the target
anode in the required size and shape → prevent bombardment of a large area on
the anode caused by the mutual repulsion of the electrons
· The
focusing cup is made of nickel.
·
Modern x-ray tubes may be supplied with a single or, more commonly, a double filament
GRID-CONTROLED
X-RAY TUBES
•
Conventional x-ray tubes contain two electrodes (cathode and anode).
•
The grid-controlled tube has the 3rd electrode → controls the flow of electrons
from the
filament
to the target.
The
third electrode is the focusing cup that surrounds the filament.
•
In conventional x-ray tubes a focusing cup is electrically connected to the
filament.
•
In the grid-controlled tube, the focusing cup is electrically negative relative
to the
filament.
the voltage across the filament grid produces an electric field along the path
of the
electron
beam → pushing the electrons even closer together.
If the voltage is large enough → the tube current may be completely pinched off
"act
like
a on & off switch for the tube current → used in cinefluorography".
5-Anode
· It
is the positive electrode inside the x-ray tube.
· The
place where the electrons hit to produce an x-ray
· 99%
of Electron interaction with anode material produces heat, as such anode has to
have a high melting point of 3370 0C.
· Tungsten
is used as an ideal anode for many reasons:
1-
High melting
point
2-
High Z number
(Z-74), which increases X-Ray production.
Anode Types:
1-stationary anode:
· It
is used in dental X-rays and portable x-ray where high power (Kv and mAs) is
not required.
· It
can’t withstand high temperatures produced by filament electron collisions.
· It
is composed of tungsten inserted into the copper block
· Copper
increases the thermal capacity and the cooling rate of the anode material
· Copper
has a low melting point
2-Rotating anode
· The
anode is rotating to increase the area of filament electron collisions on the
target anode.
· This
causes a good heat distribution (increases the heat capacity ) to withstand the
tube heat loading.
· The
rotating anode is used in radiographic X-ray, CT, and cardiac imaging because
of the high parameters KV,mAs,and power.
·
Most rotating
anodes revolve at 3400 rpm (revolutions per minute). • The anodes of
high-capacity x-ray tubes rotate at 10,000 rpm.
·
The stem of the
anode is the shaft between the anode and the rotor.
·
It is narrow so
as to reduce its thermal conductivity.
·
The stem usually
is made of molybdenum because molybdenum is a poor heat conductor.
·
Occasionally,
the rotor mechanism of a rotating anode tube fails. • When this happens, the
anode becomes overheated and pits or cracks, causing tube failure.
A
stationary anode tube with a 1-mm focal spot may have a target area of 4 mm2.
A 15-cm diameter rotating anode tube can have a target area of approximately
1800 mm2, which increases the heating capacity of the tube by a
factor of nearly 500.
3-
Line
focus principle
• Not all of the anode is involved in
x-ray production. The radiation is produced in a very small area on the surface
of the anode known as the focal spot.
· The
dimensions of the focal spot are determined by the dimensions of the electron
beam arriving from the cathode.
· In most x-ray tubes, the focal spot is
approximately rectangular. The dimensions of focal spots usually range from 0.1
mm to 2 mm.
· X-ray
tubes are designed to have specific focal spot sizes; small focal
spots produce less blurring and better visibility of detail, and large
focal spots have a greater heat-dissipating capacity.
· Focal
spot size is one factor that must be considered when selecting an x-ray tube
for a specific application. Tubes with small focal spots are used when high
image visibility of detail is essential and the amount of radiation needed is
relatively low because of small and thin body regions as in mammography.
Most x-ray tubes have two focal spot sizes (small and large), which can be
selected by the operator according to the imaging procedure.
· Radiology
requires small focal spots because the smaller the focal spot, the better the
spatial resolution of the image.
• Unfortunately, as the size of the
focal spot decreases, the heating of the target is concentrated onto a smaller
area.
·
Before the
rotating anode was developed, another design was incorporated into x-ray tube
targets to allow a large area of heating while maintaining a small focal spot.
This design is known as the line-focus principle.
· By angling the target(Angle θ → the angle
between the central ray and the target face)
, one makes the effective area of the target much smaller than the actual area of electron interaction.
·
The line-focus
principle allows high anode heating with small effective focal spots. As the
target angle decreases, so too does the effective focal spot size.
·
The effective
target area, or effective focal spot size, is the area projected onto the
patient and the image receptor.
·
When the target
angle is made smaller, the effective focal spot size also is made smaller.
·
Diagnostic x-ray
tubes have target angles that vary from approximately 5 to 20 degrees
· Field
coverage and effective focal spot length vary with the anode angle. A. A
large anode angle provides good field coverage at a given distance; however, to
achieve a small effective focal spot, a small actual focal area limits power
loading. B. A large anode angle provides good field coverage,
and achievement of high-power loading requires a large focal area; however,
geometric blurring and image degradation occur. C. A small
anode angle limits field coverage at a given distance; however, a small
effective focal spot is achieved with a large focal area for high power loading.
· Effective
focal spot size is 0.3mm for mammography while reaching to 1.2mm for
radiography.
· Smaller
focal spot size is mandatory for better resolution, and small field coverage,
but increases the tube loading (heat loading).
· Effective
focal spot size for different imaging modalities:
· Within
the design of an x-ray tube there is a balance between the dual aims of
increasing the focal spot size as far as possible and yet maximizing heat
dissipation. There are two features that work towards optimizing these two
antagonists.
· Using
a rotating disc anode (instead of a stationary one) to spread the heat out over
the circumference of the disc. Rotating discs are usually not possible for
dental radiography, mobile units, and intra-operative radiography owing to the
smaller size of the set-up used. Stationary anodes are still used for most of
these.
· Changing
the target angle - the larger the angle the larger the effective spot size but
the worse the heat loading due to a smaller actual spot size.
Large focal spot size is useful because it means a larger area of the patient
can be captured in the resultant field of view (remembering that the photons
diverge from their point of origin although diagrammatically represented in
parallel in this diagram), but the usefulness of a large focal spot also needs
to be balanced against the degree of geometrical unsharpness it produces.
The design of the anode also helps reduced heat loading further:
A larger disc circumference means a greater area for the heat to spread over.
· Disc sizes of 20-30cm are usual
The disc itself (into which sits a ring of target material) can be made from molybdenum
which has a high melting point and low density but is a poor conductor of heat,
with a thin molybdenum stem transmitting the rotation from the motorized part
of the assembly.
· The
target material itself can be made from an alloy of tungsten and rhenium which
does not coarsen as fast as tungsten on its own.
4-Anode
heel effect
· The
x-ray photons are produced at the focal spot area to travel in a circular path.
· The
only photons moving towards the window of the X-ray tube can reach the patient.
· The
photons move a certain distance through the target anode before leaving it.
· The
angulation of the target anode causes unequal distance traveled by photons from
the interaction site to the surface of the target material.
·
· The
distance traveled is long for photons moving toward the anode, while short for
photons moving toward the cathode.
· The
target anode material cause attenuation to the x-ray photons which reduces the
intensity of x-ray photons on the anode side more than on the cathode side.
·
· The
x-ray beam becomes not uniform, where the x-ray photons have a higher intensity
at the cathode side than photons at the anode side.
· Smaller
anode angle causes an increase in the distance traveled by photons moving
towards the anode side and subsequently reduces the X-Ray intensity at the
anode side.
·
Factors
affecting the healing effect:
· Anode Angle: reducing the anode
angle causes an increase in the heel effect
· Small
anode angle
·
Large
anode angle
·
Field
size:
· Using
less field size(small coverage area ) by tight collimation cause fewer
variations in X-Ray photon intensities
· Large
field size causes high variation between X-ray photons intensities at the cathode
and anode sides as shown below:
·
·
Less
field size cause less variation in X Ray photon intensities
·
Source
to image receptor distance SID
· The
distance from the X-ray origin to the image receptor at the anode side is
shorter than the distance towards the cathode side.
· If
SID increases the variation in X-ray intensities decreases at the anode
direction against the central line.
· If
SID increases the variation in X-ray intensities decreases at the cathode
direction against the central line.
·
· the
variation in X-ray intensities decrease at the anode direction against the
central line
·
·
the
variation in X-ray intensities decrease
at the cathode direction against
the central line
·
· If
SID increases, the loss of X-ray relative intensity at the cathode side is
higher than at the anode side
Uses
of heel effect :
Chest
X Ray imaging :
Make the cathode side
towards the abdomen region, as the abdomen is high attenuating organ than the lung
Abdomen
X-ray imaging :
•
Cathode side is
directed toward the upper abdomen where high attenuation happens to the X-Ray
beam.
•
Anode side is
directed towards the lower abdomen where lower X-Ray attenuation happen
•
Mammography
imaging
•
Positioning the
cathode toward the thickest part (Chest wall) for obtaining uniform image .
•
5-off
focus radiation:
· Off-focus
radiation results from rebounding electrons from the focal spot striking other
areas of the anode, thereby producing a large low-intensity x-ray source.
· This
extra-focal radiation increases the patient dose and image blurring from
shadowing and decreases image contrast.
· Patient anatomy appearing outside of the
exposure field (e.g., ears on a skull examination) is attributed to off-focus
radiation.
· The upper fixed shutters of a collimator
(beam-limiting device) help reduce the amount of off-focus radiation reaching
the film.
6-Tube
rating and cooling
· The
heat loading (rating) depends on acquisition parameters Kv., mA,…..
· When
Kv or mA increase, the number of filament electrons increases which leads to
higher collisions and heat dissipation.
· Although
the joule is the basic unit for energy and heat, it is not always used to
express x-ray tube heat. The special heat unit (HU) was
introduced when single-phase equipment was common to make it easy to calculate
heat.
·
The relationship between a quantity of heat expressed in heat units and
in joules is given by
· Heat (HU) = 1.4 x heat (J).
· Note:
A heating unit is a smaller quantity of heat than a joule since one joule is
equal to 1.4 heat units.
·
Since the product of the joules-to-heat unit conversion factor (1.4) and the
waveform factor for single-phase (0.71 ) is equal to 1, the following
relationship is obtained:
Heat (HU) = KVp x MAS.
·
Here it is seen that for single-phase operation, the heat produced in heat
units is the product of the KVp and MAS. In fact, this is why
the heat unit is used. In the earlier days of radiology, when most equipment
was single-phase, it was desirable to calculate heat quantities without having
to use a waveform factor. This was achieved by introducing a new unit, the heat
unit. For three-phase, six-pulse equipment, the heat in heat units is given by
Heat (HU) = 1.35 x KVp x MAS.
· The
factor of 1.35 is the ratio of the waveform factors, 0.96/0.71.
· The rate at
which heat is produced in a tube is equivalent to the electrical power and is
given by
Power (watts) = w x KVp x MA.
· The
total heat delivered during exposure, in joules or watt-seconds, is the product
of the power and the exposure time.
· Heat
is normally removed from the anode through radiation through the vacuum and
into the conducting oil outside the glass envelope.
· The
molybdenum stem conducts very little heat to prevent damage to the metal
bearings.
· In
x-ray tube operation, the goal is never to exceed specific critical
temperatures that produce damage.
· This is achieved by keeping the heat content
below specified critical values related to the tube's heat capacity.
· In
most x-ray tubes there are three distinct areas with critical heat capacities,
as shown below.
· The
area with the smallest capacity is the focal spot area, or track, and is the
point at which heat is produced within the tube. From this area, the heat moves
by conduction throughout the anode body and by radiation to the tube housing;
heat is also transferred, by radiation, from the anode body to the tube
housing.
· Heat is removed from the tube housing by
transfer to the surrounding atmosphere. When the tube is in operation, heat
generally flows into and out of the three areas shown. Damage can occur if the
heat content of any area exceeds its maximum heat capacity.
· it
is seen that the safe power limit of a tube is inversely related to the
exposure time. This is not surprising, since the total heat developed during
exposure is the product of power and exposure time. It is not only the total
amount of heat delivered to the tube that is crucial, but also the time in
which it is delivered.
X-ray
tube rating:
Thermal
rating
• X-ray tubes are inherently inefficient; the vast majority
(>99%) of energy used in the process becomes unwanted heat with barely 1%
being converted to x-rays. The unwanted heat can damage the x-ray tube so x-ray
machines have a rating, above which they should not be operated, in order to
reduce the chance of damaging the machine.
The parameters which
can be changed to preserve the tube rating are:
•
kV
•
mA
•
exposure time
•
Taking all these three
into consideration, each machine has a limit measured in kW per 0.1 second
within which the damage risk to the x-ray tube during use is acceptable.
• There are ways of mitigating some of the heat production in
order to try and improve the length of time that the tube can be used during
each examination before it must be switched off and allowed to cool. These
include:
•
Larger anode disc to
spread the heat over a larger area
•
Larger effective focal
spot size
•
Larger target angle
(with the same effective focal spot size)
•
Faster disc rotation
•
Since these factors
cannot be changed once the machine has been produced, in practice the thermal
rating is used as a time limitation for a given kV/mA setting. While setting up
an image, the kVs and mAs can be adjusted (or are set to protocols) to give a
time length of how long the Xray exposure can be without causing heat stress on
the machine.
The
maximum allowable power decreases while Tube loading , Heat loading
increase with:
• Lengthening
exposure time
• Decreasing
effective focal spot size (heat spread over a smaller area)
This
means the heat is spread over a smaller area and the
rate of heat dissipation is reduced)
• Decreasing
disk diameter (heat spread over smaller circumference and area)
• Decreasing
the speed of disk rotation
• There
is more efficient heat loss from a rotating anode, so the rating is higher.
• A
larger focal spot causes less heating than if the beam were focused onto a
smaller
area.
• A
smaller anode angle has a higher heat rating.
• Rating
is increased with full wave rectification.
Other
factors to take into consideration are:
• By
using a higher mA, the maximum kV is reduced, and vice versa.
• A
very short examination may require a higher power to produce an adequate image.
This must be taken into consideration as the tube may not be able to cope with
that amount of heat production over such a short period of time.
Anode
cooling chart
• As
well as withstanding high temperatures an anode must be able to release the
heat quickly too. This ability is represented in the anode cooling chart. It
shows how long it takes for the anode to cool down from its maximum level of
heat and is used to prevent damage to the anode by giving sufficient time to
cool between exposures.
•
Others
• Window:
made of beryllium with aluminum or copper to filter out the soft x-rays. Softer
(lower energy) x-ray photons contribute to patient dose but not to image
production as they do not have enough energy to pass through the patient to the
detector. Therefore, to reduce this redundant radiation dose
• To
the patient, these x-ray photons are removed.
• Insulating
oil: carries heat produced by the anode away via conduction.
• Filter:
Total filtration must be >2.5 mm aluminum equivalent (meaning that
the
material provides the same amount of filtration as a >2.5 mm thickness of
aluminum)
for a >110 kV generator
Total filtration =
inherent filtration + additional filtration
(removable
filter)
7-Sample Questions
How much power does a
typical chest radiograph require?
Protocol: 110kVp, 32mA, 0.25s
880
Watts
3520
Watts
0.9
Watts
880,000
Watts
Which of the following
methods is not used to prevent the anode from melting in an x-ray tube?
An
oil bath
Use
of very high melting point materials
Rotating
the anode
Water/glycol
coolant
What is the
approximate mean energy of a 140kVp photon beam?
110
80
Insufficient
information provided
50
A real x-ray beam can
be characterized by the energy of a monoenergetic beam that would give the same
HVL. What is the term for this characterization?
Equivalent
KERMA
Mean
energy
Effective
energy
kVp
corrected energy
What is the
approximate efficiency of a 120kVp x-ray tube using a Tungsten target?
27.4%
9.7%
0.8%
3.4%
How much energy does a
typical chest radiograph require?
Protocol: 110kVp, 32mA, 0.25s
3520
Joules
880
Joules
0.9
Joules
880,000
Joules
Concerning properties of x-rays:
A. Beam intensity is the total energy per unit area per unit
time.
B. The inverse square law applies to all x-ray beams
C. X-rays have lower linear energy transfer than alpha
particles.
D. All electromagnetic radiation can cause ionization.
E. At equivalent
energy, an x-ray cannot be distinguished from a gamma ray.
Concerning an x-ray tube:
A. Usually a voltage of 10V and a current of 10A pass through
the filament.
B. The accelerating voltage of the tube is typically in the
range 60-120k\/.
C. The process of thermionic emission occurs on the surface of
the anode.
D. When an accelerating electron interacts closely with a target
nucleus it is deflected and
slowed, losing energy that is emitted as an x-ray photon.
E. The angle of the target ensures that all x-rays produced pass
through the window in the tube to form a beam.
X-ray production in a diagnostic x-ray tube:
A. Occurs when moving electrons interact with target nuclei.
B. is 99% efficient.
C. Is more efficient with a rotating compared to a stationary
anode.
D. Is increased if the target atomic number is increased.
E. Requires a cooling air current at all times within the tube.
Radiation output from an x-ray tube increases with:
A. Increasing cathode-to-anode distance.
B. The addition of a filter.
C. Increasing kV.
D. Increasing mA.
E. A constant potential compared to a single-phase waveform.
In a diagnostic x-ray tube:
A Heat is only removed to the tube envelope by conduction.
B. The focusing cup is negatively charged.
C. Rotor bearings are lubricated with oil.
D. The anode stem is a poor heat conductor.
E. The addition of rhenium to the tungsten target increases the
toughness and lifespan of the target.
A False: Heat is radiated through
the vacuum to the envelope.
B. True: To repel electrons from
the filament and produce a narrow beam.
C. False: Silver is used.oil
would evaporate in the vacuum.
D. True: Prevents damage to the
rotor assembly.
E. True: This alloy reduces surface
pitting and increases lifespan.
In a diagnostic x-ray tube, the
anode angle:
A. Is the angle that the target
face makes with the x-ray beam.
B. Is the only factor determining
focal spot size.
C. Is generally 20-35°.
D. Increases the tube rating if
the angle is reduced.
E. Determines the size of field
covered by the x-ray beam at a given focus-film distance.
The spectrum of an x-ray beam:
A. Is not affected by filtration.
B. Varies with tube current.
C. Has a maximum energy
determined by peak tube potential (kVp).
D. Consists of Bremsstrahlung and
characteristic radiation if the kVp exceeds the K edge
energy of the anode.
E. Has a peak approximately % of
the kVp.
Regarding an x-ray tube filament:
A. It must have a high melting
point and low resistance.
B. Electrons evaporate off
through thermionic emission.
C. Tungsten is used because of
its high atomic number:
D. It should have a low vapour
pressure.
E. It has a negative potential.
The anode-heel effect:
A. Produces a uniform x-ray beam
across its field.
B. Is more prominent at the
cathode end of the tube.
C. Is more noticeable if the
focus-film distance is increased.
D. Is greater if the target angle
is steeper:
E. Is not used as an advantage in
diagnostic radiography.
' Concerning attenuation of
x-rays:
A. Increased tube filtration
increases the half-value layer:
B. Total attenuation is the
product of Compton, photoelectric, and elastic attenuation
effects.
C. Half-value thickness is
inversely proportional to the linear attenuation coefficient.
D. It is altered with differing
atomic number materials.
E. It is related to the inverse
square law.
X-ray tube rating increases with:
A. Rotating compared to
stationary anodes.
B. Larger focal spot size.
C. Increasing the anode angle
with fixed focal spot size.
D. Half wave compared to full
wave rectification.
E. Quicker production of heat.
The half-value layer:
A Is the thickness of a material
that will reduce the intensity of a narrow x-ray beam to Y.
of its original value.
B. Is a measure of the
penetrating power of an x-ray beam.
C. For lead is greater than for
aluminum at a given energy of the x-ray beam.
D. Is reduced as the photon
energy of the radiation decreases.
E. Will produce exponential
attenuation if a narrow x-ray beam passes through successive
Half-value thicknesses of a
particular material
Comments
Post a Comment